Public comment period · §3

Draft standard CEI-MS-04: Determination of water content in… — disposition

The disposition of every submission and the amendments that resulted.

Document identifier
CEI-CP-086/3
Series
Public comment period
Version
1.0
Published
05 Jun 2026
Last reviewed
05 Jun 2026
Next review
05 Jun 2027
Identifier
10.71829/cei.cp.86
Certainty
Not rated
Cycle
2026 Q2
Window
09 Apr 2026 – 09 May 2026
Status
Closed
Submissions
16

§3Disposition and amendments

§3.1Disposition table

Table 1. Every submission with its disposition. Each row links to the submission in full.

ReferenceRespondentPoint raisedDisposition
DRAFT-WATER-KARL-F/001Dr Melisande Thorsby-NakamuraNothing in the standard addresses transfer of the method between laboratoriesAccepted in part
DRAFT-WATER-KARL-F/002Dr Vasilisa ImmelmannNothing prevents a chromatogram being reprocessed until it passesAccepted in part
DRAFT-WATER-KARL-F/003Dr Lorcan UttridgeThe standard should require documented operator competenceNot accepted
DRAFT-WATER-KARL-F/004Dr Rurik Haverkamp-DialloDeclared interests should appear on the document rather than on a separate pageNoted, no amendment
DRAFT-WATER-KARL-F/005Rukayat Zaleski-MbekiNothing requires a laboratory to demonstrate ongoing performanceAccepted in part
DRAFT-WATER-KARL-F/006Quentin Whitmarsh-ObiA superseded version should remain reachable from the version that replaced itNoted, no amendment
DRAFT-WATER-KARL-F/007Dr Sigrún VercingetorixAcceptance criteria are expressed in terms specific to one instrument familyAccepted in part
DRAFT-WATER-KARL-F/008Dr Kamila Glendinning-UcheThe standard assumes a reference standard of known assigned contentAccepted
DRAFT-WATER-KARL-F/009Dr Fitzwilliam Danquah-ÖbergThe draft prescribes a procedure where it should state performanceAccepted
DRAFT-WATER-KARL-F/010Dr Evander Whitmarsh-ObiA system suitability criterion is unachievable on instrumentation in general useAccepted in part
DRAFT-WATER-KARL-F/011Dr Quentin ZimmerthalNothing states how long a prepared standard solution remains usableAccepted
DRAFT-WATER-KARL-F/012Dr Eamon ImmelmannQuantitative claims are reproduced without the method that produced themAccepted
DRAFT-WATER-KARL-F/013Dr Anselm Thorsby-NakamuraResults are reported without an estimate of measurement uncertaintyAccepted in part
DRAFT-WATER-KARL-F/014Dr Xenia Nyquist-ObioraThe standard should state a pass limitNot accepted
DRAFT-WATER-KARL-F/015Dr Brigitta Grünbaum-SowandeA single method cannot support the conclusion the standard permitsAccepted in part
DRAFT-WATER-KARL-F/016Dr Yehudit YorkstoneThe preparation solvent is not specified and changes what is measuredAccepted in part
16 submissions in total.

§3.2Summary by disposition

Table 2. Counts by disposition, with the meaning of each.

DispositionCountMeaning
Accepted4The submission is accepted and the draft is amended as proposed. The amendment is recorded in the amendment log of the document it changed and is traceable to the numbered submission.
Accepted in part8Part of the submission is accepted and part is not. The secretariat response states which part is which and on what ground the remainder was declined.
Noted, no amendment2The submission raises a point the Institute accepts but that does not require a change to the draft, most often because the draft already states it elsewhere. The location is given in the response.
Not accepted2The submission is not accepted. The secretariat response gives the reason. A submission that is not accepted remains published in full; the Institute does not remove a submission because it disagrees with it.

§3.3Resulting amendments

  1. Nothing in the standard addresses transfer of the method between laboratories — arising from DRAFT-WATER-KARL-F/001. The standard now states the maximum difference between laboratories analysing the same sample at which results are treated as agreeing, and requires that a comparison across laboratories state whether that criterion was met.
  2. Nothing prevents a chromatogram being reprocessed until it passes — arising from DRAFT-WATER-KARL-F/002. The standard now requires the processing method and its version to be recorded with every result and the primary data to be retained. A full audit trail is recommended and not required, since it depends on software many laboratories in scope do not run.
  3. Nothing requires a laboratory to demonstrate ongoing performance — arising from DRAFT-WATER-KARL-F/005. The standard now requires that system suitability results be trended across runs and that a determination report whether the trend was within the laboratory's stated limits, which places the demonstration inside the document the Institute can read.
  4. Acceptance criteria are expressed in terms specific to one instrument family — arising from DRAFT-WATER-KARL-F/007. The acceptance criteria are now expressed in instrument-independent quantities, with the conversion from the common manufacturer-specific reporting units given in a note so that a laboratory can check conformance without recalculating by hand.
  5. The standard assumes a reference standard of known assigned content — arising from DRAFT-WATER-KARL-F/008. The standard now states which determinations remain available without a reference standard of assigned content, requires that such determinations be labelled as relative rather than absolute, and requires the assumption used in place of an assigned content to…
  6. The draft prescribes a procedure where it should state performance — arising from DRAFT-WATER-KARL-F/009. The prescriptive clauses are converted to performance requirements, the specific conditions are retained as a worked example clearly labelled as one, and a laboratory using a different validated procedure that meets the stated performance reports a conformant…
  7. A system suitability criterion is unachievable on instrumentation in general use — arising from DRAFT-WATER-KARL-F/010. The suitability criterion is now stated at two levels, a general level achievable on instrumentation in common use and a stricter level required only where the standard is applied to the discrimination named in the scope, with the reason for the difference…
  8. Nothing states how long a prepared standard solution remains usable — arising from DRAFT-WATER-KARL-F/011. The procedure now requires solution stability to be demonstrated over the period of intended use, and the demonstration to be recorded with the result.
  9. Quantitative claims are reproduced without the method that produced them — arising from DRAFT-WATER-KARL-F/012. Every quantitative claim now carries the determination that produced it at the point of use, and figures obtained by non-comparable methods are no longer presented in the same row or sentence.
  10. Results are reported without an estimate of measurement uncertainty — arising from DRAFT-WATER-KARL-F/013. Every determination now reports the repeatability observed in the run, the reporting precision, and a statement that two results differing by less than the stated repeatability are not distinguishable by this method.
  11. A single method cannot support the conclusion the standard permits — arising from DRAFT-WATER-KARL-F/015. The standard now distinguishes a determination of the measured quantity, which stands alone, from a conclusion about the material, which requires a named orthogonal determination, and states that the second claim made on the strength of the first alone is not…
  12. The preparation solvent is not specified and changes what is measured — arising from DRAFT-WATER-KARL-F/016. The preparation solvent is now a required reporting field and the procedure names an acceptable range. Specifying one solvent for every material in scope would exclude compounds the standard is intended to cover.

References cited on this page

References are numbered in order of first citation in this document. Each superscript in the text links to its entry below.

  1. International Organization for Standardization. ISO/IEC 17025:2017 General Requirements for the Competence of Testing and Calibration Laboratories. ISO/IEC Standard 2017;3rd edition. identifier not held by the Institute

Identifiers are reproduced only where the Institute holds them. Where a digital object identifier or PubMed identifier is not shown, the Institute has recorded the journal and year and has not constructed an identifier.

Nothing published by the Institute is medical advice, a diagnosis, a prescription, a treatment recommendation or a purchasing recommendation. Compounds supplied for research use are not approved for human or veterinary use in any jurisdiction, and a favourable analytical assessment of a supplier is not a statement that any product is safe or effective. The Institute publishes certainty ratings and never recommendations. No telephone number, messaging handle or ordering channel appears anywhere on this site.